ERROR RVI Precision Station : Data Error (PM))

26 January 2024 - 05 February 2024
26/01/2024 : 11:00 The RVI Precision Station1135100 is showing a Device Status issue on the Sensor Flow :
 
The flow volume is about 4.8l/min which looks normal. However the flow velocity is about 7.5 : it is usually around 9 to 10 m/s.
 
Spoke to Acoem on Friday 26th, unusual to have an issue with the flow velocity (as opposed to flow volume) and was suggested to re-booting and then seeing if it cured it when it settled down. Did this : visited this morning but still showing the warning afterwards and a coupleof days later.
 

31/01/2024 Callout Requested

05/02/2024 12:41 - 14:45 Acoem Engineer on site.

See Engineers Report

05_02_2024_Newcastle University_RVI_Callout Instrument Fault_11943 & 12283_Marius Barbulescu.pdf

Resolved

Data may not be accurate during this period.

 