ERROR COAST RD Precision Station : Data Error (All gas analysers))
10 August 2022 - 07 September 2022
Because of issues with intermittent Air Conditioning at both these sites, the gas analysers were showing high Internal and Cooler temperature related alarms.
there is a risk of damage to the gas analysers at these sites during the next few days of hot weather.
To try and protect the analysers from potential damage due to overheating during the next few days of hot weather, the gas analysers have been shut down for the next few days until the weather cools down a bit.
06/09/22 - 07/09/2022 Air-conditioning Unit removed back to AC company, deep cleaned.
Data will be missing during this period.