ERROR GOSFORTH Precision AQ : BAD DATA (NO2, Nox, NO)

16 April 2022 - 20 May 2022

16/04/2022 10:00 Nitrogen readings spiked high.

20/04/2022 Investigation shows Nitrogen Analyer has a Cooler Temperature High alarm (Actual 99.9, Max -1.0).

21/04/2022 Requested callout.

25/04/2022 10:00 Acoem Engineer diagnosed Nitrogen Analyser PMT Cooler needs replacement. Turned off analyser.

See Engineer report 25_04_2022_Newcastle University_Gosforth _Routine Service_Callout_AML10262-21-01_Marius Barbulescu.pdf

19/05/2022 14:00 Swapped the Nitrogen Analyser at Four Lane Ends (1155100) with the faulty Gosforth analyser.

20/05/2022 10:45 Some local Mobile Data network issues plus difference in baud rate (changed new Gosforth Nox to work at 11520 baud for consistency at Gosforth Logger) meant this did not start reporting Nox till now.

Data between this period should either be oignored or will be missing.

,

 