GOSFORTH Precision AQ : ERROR (CO)
19/05/2020 CO data looks unreliable after the service - spiking negative.
08/06/2020 Reported to Airmonitors.
09/06/2020 Airmonitors engineer called out. Adjusted SR ratio.
For Engineers report, see Urban Observatory Github
12/06/2020 Re-occurrence. Reported
15/05/2020 12:15 - 15:00 2nd Callout, for Engineers Reports see Urban Observatory Github
15/05/2020 21:00 Still not fixed- now reading massively high from a few hours after engineer left.
17/06/2020 Reported again, requesting uplift for repair.
22/06/2020 :12:05 CO Analyzer removed for repair.
TODO Awaiting return
27/08/2020 13:25 Engineer reinstated CO analyser. Analyser Serial Number is 1161530004, DA200620.
28/08/2020 16:15 Engineer did final checks.
For more details see Engineers Report in Urban Observatory Github: