ANGEL Precision AQ : CO Disconnected
23 March 2022
23/03/22 15:15 Decomissioned CO Analyser (Serial No 1180920066, Model Code 48i-BCSAA) as service contract not renewed and Ricardo say it has been showing abnormal and excessively high values .
CO Analyser may be faulty : 2 alarms being shown :
Motor Speed Low Alarm : 25% (should be near 100%)
AGC Intensity Low Alarm : actual 7135Hz (Minimum is 150000)
CO Data will not be collected from this date.